Double-layer glass reactor common problems and solutions

Double-layer glass reactor common problems and solutions

Clicks:687 Release time:2010/5/21

Now    Elephant
original    because
row    except
Turn on the power switch and the indicator light does not light.
External power supply is not connected or has poor contact
Ask a professional electrician to check the power supply and socket.
Fuse short circuit
Set the power switch to the OFF position and replace the fuse.
The power indicator is on but not rotating.
Rotating shaft rust
Stop using, with the supplier
Motor, electrical box failure
7-pin plug not connected
Reconnect the seven-pin plug
The vacuum suddenly disappears
There is crack in the glass and the switch is broken.
Check the glass parts and switch the switch
There is a vacuum, but I can’t get it.
The seal is worn and the vacuum switch is connected to leak
Please replace the seal, switch
Sometimes there is no vacuum
There is a loose connector on the steel shaft.
Please remove the dirt check vacuum gauge, vacuum pump
Vacuum hose aging
Please replace the vacuum hose
The main mandrel is loose left and right
Tightening screw
Screw tight
Motor temperature is too high (room temperature plus 40 °C is normal)
overload
Stop, turn the shaft by hand, whether it is heavy, remove the dirt on the contact part of the sealing ring and the glass shaft, apply vacuum grease
The speed display does not match the actual
The voltage is unstable and has its own error
With suppliers
Enclosure charged
The heating pipe has cracks and water
Please check with a professional electrician