Bench-top constant temperature culture shaker operating instructions:

1. Plug in the power cord, turn on the power switch, open the lid, place the container in the appropriate fixture, and then close the lid.

2. Turn on the corresponding oscillation switch and adjust the speed to the required speed. Since the machine has two sets of oscillation systems, the two sets of system switches must be opened separately. The two systems can be used simultaneously or separately.

3. If you need to control the temperature, you can turn on the temperature control switch and set the required temperature on the temperature controller. The operation instructions of the temperature controller of the machine can be found in the operation manual of the temperature controller.

4. When using, first adjust the oscillation speed to zoi low and turn off each group of oscillation switch and temperature control switch, then turn off the main power supply.

5. The local timer is generally set to the ON position (ie, normally open). If you need to use the timing function, you can adjust the timer to the desired position. When the timer is finished, the unit is in the off state.

Desktop constant temperature culture shaker troubleshooting:

1. If the unit cannot be started after turning on the power, check whether the power switch indicator is on. If the power indicator is on, check whether each group of oscillation switches is turned on and the governor is working. If the power indicator is off, check it. Is the fuse behind the unit intact and the plug plugged in?

2. If you can't control the temperature, you can check if the temperature control switch is turned on. If the temperature control switch is turned on, check if there is a digital display on the temperature controller and if the temperature control setting is correct.

3. If it is impossible to illuminate, check if the light switch has been turned on and the lamp is broken. If the lamp is broken, just replace it with a new one.

4. If you are unable to find out the problem, please contact our company directly, and do not open the instrument by yourself.